FDA recall Z-1340-2022

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444896/RF521

Reason for recall

Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.

Key facts

Status
Terminated
Initiation date
2022-05-18
Report date
2022-07-13
Termination date
2025-02-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newark, DE, United States