FDA recall Z-1341-2020

Hitachi Medical Systems America Inc · Class II · device

Product

UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel.

Reason for recall

These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.

Distribution

Domestic distribution to CA, FL, GA, MA, NV, OH, TX, and WV. Foreign distribution to Canada.

Key facts

Status
Terminated
Initiation date
2020-01-31
Report date
2020-02-26
Termination date
2020-07-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Twinsburg, OH, United States