FDA recall Z-1341-2021

Heartware, Inc. · Class I · device

Product

Medtronic HVAD Packaged Monitor, Model Numbers: 1500, 1500AU, 1500DE, 1510US, 1511AR, 1511AU, 1511BR, 1511CA, 1511CH, 1511DE, 1511GB, 1511IL, 1511IN, 1511IT, 1511JP, 1511MY, 1520CLIN-AU, 1520CLIN-DE, 1520CLIN-GB, 1520JP, 1520US, 1521AU, 1521BR, 1521CA, 1521CH, 1521DE, 1521GB, 1521IL, 1521IN, 1521IT, 1521JP, 1521KR, 1521TW, 1521US, 1522CA, 1522CA-CLIN

Reason for recall

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Distribution

Worldwide

Key facts

Status
Ongoing
Initiation date
2021-02-26
Report date
2021-04-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Miami Lakes, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1341-2021