FDA recall Z-1344-2024

Angiodynamics, Inc. · Class II · device

Product

MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-767

Reason for recall

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Key facts

Status
Ongoing
Initiation date
2024-01-30
Report date
2024-03-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Queensbury, NY, United States

Primary source