FDA recall Z-1350-2023

Church & Dwight Inc · Class II · device

Product

Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)

Reason for recall

Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2023-01-25
Report date
2023-04-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Princeton, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1350-2023