FDA recall Z-1352-2019

Moller Vital · Class II · device

Product

Molift 2-Point sling bar, an accessory for Molift Mover 180/Air. Item nos. 1830001 1830002 1830002C 1830003

Reason for recall

The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.

Distribution

Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.

Key facts

Status
Ongoing
Initiation date
2019-02-04
Report date
2019-05-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Gjovik, Norway

Primary source