FDA recall Z-1355-2020

Cook Inc. · Class II · device

Product

Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.

Reason for recall

The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.

Distribution

Foreign distribution only. Distributed worldwide.

Key facts

Status
Terminated
Initiation date
2020-01-16
Report date
2020-03-04
Termination date
2020-11-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1355-2020