FDA recall Z-1355-2026

Intuitive Surgical, Inc. · Class II · device

Product

da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Reason for recall

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Distribution

US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.

Key facts

Status
Ongoing
Initiation date
2025-12-30
Report date
2026-02-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sunnyvale, CA, United States