FDA recall Z-1364-2022

Fresenius Medical Care Holdings, Inc. · Class II · device

Product

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Reason for recall

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

Distribution

US Nationwide distribution in the states of NC and WV.

Key facts

Status
Terminated
Initiation date
2021-08-03
Report date
2022-07-20
Termination date
2024-04-10
Voluntary/Mandated
Voluntary: Firm initiated