FDA recall Z-1365-2020

BIOMERIEUX - Grenoble · Class II · device

Product

NUCLISENS¿ Lysis Buffer, IVD, REF 200292, 48X2 ML LYS Manufacturer bioMerieux SA 376, Chemin de l'Orme 69280 Marcy-l'Etoile - France - Product Usage: to be used for the release of total nucleic acid from biological specimens.

Reason for recall

Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction.

Distribution

No US distribution. No Mexican consignees. International distribution to South Africa, Uruguay, Tanzania, Romania, Netherlands, Macao, Italy, India, Hong Kong, United Kingdom, France, Spain, Germany, China, Chile, Canada, Belgium, Australia, Argentina, Curacao.

Key facts

Status
Terminated
Initiation date
2020-01-10
Report date
2020-03-04
Termination date
2022-04-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Grenoble Cedex 1, N/A, France