> **Draeger Medical Systems, Inc.** · Class II · device recall initiated 2022-05-20.

## Product

Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

## Reason for recall

The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.

## Distribution

No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.

## Key facts

- **Recall number:** Z-1365-2022
- **Recalling firm:** Draeger Medical Systems, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2022-05-20
- **Report date:** 2022-07-20

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Telford, PA, United States