FDA recall Z-1372-2019

Polymer Technology Systems, Inc. · Class II · device

Product

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Reason for recall

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Distribution

Domestic distribution nationwide.

Key facts

Status
Terminated
Initiation date
2019-01-16
Report date
2019-05-29
Termination date
2020-06-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1372-2019