FDA recall Z-1372-2021

Beckman Coulter Inc. · Class II · device

Product

MicroScan WalkAway-40 plus Reconditioned Instrument

Reason for recall

Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

Distribution

U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.

Key facts

Status
Terminated
Initiation date
2019-09-04
Report date
2021-04-14
Termination date
2023-01-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States