# FDA recall Z-1372-2024

> **Ortho-Clinical Diagnostics** · Class II · device recall initiated 2024-02-27.

## Product

VITROS Performance Verifier I

## Reason for recall

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

## Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

## Key facts

- **Recall number:** Z-1372-2024
- **Recalling firm:** Ortho-Clinical Diagnostics
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-02-27
- **Report date:** 2024-04-03

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Rochester, NY, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1372-2024

## Citation

> AI Analytics. FDA recall Z-1372-2024. Retrieved 2026-06-12 from https://api.ai-analytics.org/recall/Z-1372-2024. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*
