FDA recall Z-1383-2021

Covidien, LP · Class II · device

Product

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

Reason for recall

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.

Key facts

Status
Terminated
Initiation date
2021-03-04
Report date
2021-04-21
Termination date
2024-07-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
North Haven, CT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1383-2021