FDA recall Z-1384-2023

Encore Medical, LP · Class II · device

Product

EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108

Reason for recall

Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.

Distribution

US: TX, IN, MD, CA, KS, AL, HI, WA

Key facts

Status
Ongoing
Initiation date
2023-02-28
Report date
2023-04-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Austin, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1384-2023