FDA recall Z-1388-2018

Life Technologies Corporation · Class II · device

Product

Human Fetal Hemoglobin (HBF-1) FITC Conjugate

Reason for recall

Three lots of Human Fetal Hemoglobin FITC Conjugate are displaying high background and unusual bright staining in the controls and patient samples which could potentially lead to erroneous results (i.e. false positives)

Distribution

MI, CA, IL, TX, SC, NY, MN, DE, NJ, WA , OR, IN, MA, DE, FL, PA

Key facts

Status
Terminated
Initiation date
2018-02-07
Report date
2018-04-25
Termination date
2019-05-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Frederick, MD, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1388-2018