FDA recall Z-1391-2026

Medline Industries, LP · Class II · device

Product

Medline Convenience Kits: 1) PICC LINE TRAY, Model Number: 00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number: ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Number: ACC010268C; 4) TRAY 6 CENTRAL LINE A, Model Number: ACC010487; 5) PEDS PICC INSERTION TRAY, Model Number: CVI3650; 6) LINE ATTIRE KIT, Model Number: DYKM2170; 7) KIT SURG ONC LAP DIAG/PORT INS, Model Number: DYKMBNDL153C; 8) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43A; 9) KIT SURG ONC PORT INSERTION, Model Number: DYKMBNDL43C; 10) VEIN PROCEDURE KIT, Model Number: DYNDA2076B; 11) VEIN PROCEDURE KIT, Model Number: DYNDA2076D; 12) VEIN PROCEDURE KIT, Model Number: DYNDA2076F; 13) PLACENTA KIT, Model Number: DYNDA2322B; 14) NICU CENTRAL LINE TRAY, Model Number: DYNDC2407A; 15) NICU CENTRAL LINE TRAY, Model Number: DYNDC2407B; 16) PICC LINE TRAY, Model Number: DYNJ40500A; 17) VENOUS ACCESS PACK, Model Number: DYNJ42694C; 18) AV FISTULA CREATION PACK-LF, Model Numb

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution

Worldwide distribution - US Nationwide.

Key facts

Status
Ongoing
Initiation date
2026-01-07
Report date
2026-02-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1391-2026