FDA recall Z-1396-2025

Calyxo, Inc. · Class I · device

Product

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Reason for recall

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.

Key facts

Status
Ongoing
Initiation date
2025-02-19
Report date
2025-04-02
Voluntary/Mandated
Voluntary: Firm initiated