FDA recall Z-1397-2023

Boston Scientific Corporation · Class II · device

Product

Habib EndoHPB Bipolar Radiofrequency (RF) Catheter 8F (2.7mm). A radiofrequency (RF) catheter which provides bipolar energy to perform partial or complete ablation of tissue in the pancreatic and biliary tracts. UPN: M00500070

Reason for recall

Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is currently provided in the HABIB EndoHPB Catheter's Instructions for Use ("IFU").

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Bulgaria, Canada, Costa Rica, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, Iraq, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Netherlands, New Zealand, Oman, Panama, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Unit. Arab Emir.

Key facts

Status
Ongoing
Initiation date
2023-02-02
Report date
2023-04-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Marlborough, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1397-2023