FDA recall Z-1399-2026

Medline Industries, LP · Class II · device

Product

Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER #5, Model Number: ACC010475; 4) DRAWER 1 NEURO CART, Model Number: ACC010542; 5) DRAWER 2 NEURO CART, Model Number: ACC010543; 6) DRAWER 4, Model Number: ACC010722; 7) MALIGNANT HYPERTHERMIA CART, Model Number: ACC010896; 8) LIVER TRANSPLANT #57 CDS-2, Model Number: CDS860063U; 9) THOROCOTOMY CDS #24-RF, Model Number: CDS860066T; 10) CYSTO, Model Number: CDS981795G; 11) MAJOR PROCEDURE CDS, Model Number: CDS982414N; 12) TOTAL KNEE CDS, Model Number: CDS982823R; 13) VASCULAR CDS-1, Model Number: CDS982895I; 14) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069F; 15) HEART CABG CDS, Model Number: CDS983376S; 16) CRANIOTOMY, Model Number: CDS983611N; 17) CUH KIDNEY TRANSPLANT CDS, Model Number: CDS983653K; 18) CUH KIDNEY TRANSPLANT CDS, Model Number: CDS983653L; 19) CUH KIDNEY TRANSPLANT CDS, Model Number

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution

Worldwide distribution - US Nationwide.

Key facts

Status
Ongoing
Initiation date
2026-01-07
Report date
2026-02-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1399-2026