FDA recall Z-1406-2021

Centinel Spine, Inc. · Class II · device

Product

Prodisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for intractable symptomatic cervical disc disease (SCDD), Part number: 09.820.055S

Reason for recall

Centinel Spine learned through five customer complaints that the product was missing the "UP" etching on the superior plate of the prodisc C device. Per the Surgical Technique, the "UP" is one of the methods used to indicate the direction in which the surgeon should place the prodisc C into intradiscal space. There are also "UP" indications in the packaging and on instrumentation as additional guides to help the surgeon properly implant the device that were present for the affected product.

Distribution

US Nationwide distribution in the states of CA, MD, DC, TN, GA, NC, PA, KY, NY, AZ, CO, OH, FL, MI.

Key facts

Status
Terminated
Initiation date
2021-03-05
Report date
2021-04-21
Termination date
2021-08-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
West Chester, PA, United States

Primary source