FDA recall Z-1408-2020

Medtronic, PLC · Class II · device

Product

McGRATH X3 Disposable Laryngoscope Blades-used with the McGRATH MAC Video Laryngoscope to aid the intubation of the trachea Code: X3-002-000 (Individual Unit); X3-003-000 (Carton of 10 Units) - Product Usage: is a tool used to aid the intubation of the trachea. As a rigid laryngoscope it holds and shapes the anatomy allowing a clear view of the larynx and entrance to the trachea.

Reason for recall

Blades do not fit the McGRATH MAC video laryngoscope handle or were tight fitting on the handles which may result in a change in optical performance as displayed on the video screen

Distribution

Worldwide distribution - US Nationwide and countries Australia, Canada, New Zealand, India, Korea, Japan, Indonesia, Malaysia, Philippines, Singapore, Vietnam, Latin America, France, Switzerland. Italy, Belgium, Germany, Netherlands, Denmark, Poland, Austria, Czech Republic, Croatia, Spain, Finland,Greece, Guadeloupe, Ireland, Hungry, Italy, Jordan, Kazakhstan, Lithuania, Lebanon, Luxembourg, Norway, Poland, Portugal, Reunion, Romania, Russian Federation, Saudia Arabia, Sweden, Serbia, Slovakia, Turkey, Tunisia, United Arab Emirates, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2020-01-28
Report date
2020-03-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
North Haven, CT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1408-2020