# FDA recall Z-1414-2026

> **Medline Industries, LP** · Class II · device recall initiated 2026-01-07.

## Product

Medline Convenience Kits:  1) OB KIT, Model Number: DYKM2326;   2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165;   3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31;   4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38;   5) L & D PACK-LF, Model Number: DYNJ0456816O;   6) L & D PACK-LF, Model Number: DYNJ0456816P;   7) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S;   8) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B;   9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C;   10) DELIVERY PACK-LF, Model Number: DYNJ20546F;   11) VAGINAL DELIVERY PK, Model Number: DYNJ20979I;   12) VAGINAL DELIVERY PK, Model Number: DYNJ20979J;   13) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D;   14) OB PACK-LF, Model Number: DYNJ26121I;   15) OB DELIVERY PACK, Model Number: DYNJ31024J;   16) OB DELIVERY PACK-LF, Model Number: DYNJ32139C;   17) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F;   18) MAJOR VAGINAL PACK, Model Number: DYNJ35488F;   19) ENS. ACCOUCHEMENT 

## Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

## Distribution

Worldwide distribution - US Nationwide.

## Key facts

- **Recall number:** Z-1414-2026
- **Recalling firm:** Medline Industries, LP
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-01-07
- **Report date:** 2026-02-25

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Northfield, IL, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1414-2026