Medline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Number: CDSCCR611P; 5) LUD C-SECTION-LF, Model Number: DYNJ0101616I; 6) C-SECTION TRAY, Model Number: DYNJ15494G; 7) C-SECTION TRAY-LF, Model Number: DYNJ19529P; 8) C-SECTION TRAY-LF, Model Number: DYNJ19529Q; 9) C-SECTION TRAY-LF, Model Number: DYNJ19529R; 10) C-SECTION TRAY-LF, Model Number: DYNJ19529S; 11) C-SECTION PACK-LF, Model Number: DYNJ21550Q; 12) C-SECTION PACK, Model Number: DYNJ23544F; 13) C-SECTION PACK, Model Number: DYNJ32632F; 14) C-SECTION PACK QVH-LF, Model Number: DYNJ35100I; 15) C-SECTION PACK-LF, Model Number: DYNJ38801F; 16) SCRIPPS SW C-SECTION PACK-LF, Model Number: DYNJ40129C; 17) C-SECTION, Model Number: DYNJ41444F; 18) C-SECTION PACK, Model Number: DYNJ49224I; 19) C SECTION PACK-LF, Model Number: DYNJ52387I; 20) C-SECTION PACK, Model Number: DYNJ56
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Worldwide distribution - US Nationwide.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1417-2026