FDA recall Z-1417-2026

Medline Industries, LP · Class II · device

Product

Medline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Number: CDSCCR611P; 5) LUD C-SECTION-LF, Model Number: DYNJ0101616I; 6) C-SECTION TRAY, Model Number: DYNJ15494G; 7) C-SECTION TRAY-LF, Model Number: DYNJ19529P; 8) C-SECTION TRAY-LF, Model Number: DYNJ19529Q; 9) C-SECTION TRAY-LF, Model Number: DYNJ19529R; 10) C-SECTION TRAY-LF, Model Number: DYNJ19529S; 11) C-SECTION PACK-LF, Model Number: DYNJ21550Q; 12) C-SECTION PACK, Model Number: DYNJ23544F; 13) C-SECTION PACK, Model Number: DYNJ32632F; 14) C-SECTION PACK QVH-LF, Model Number: DYNJ35100I; 15) C-SECTION PACK-LF, Model Number: DYNJ38801F; 16) SCRIPPS SW C-SECTION PACK-LF, Model Number: DYNJ40129C; 17) C-SECTION, Model Number: DYNJ41444F; 18) C-SECTION PACK, Model Number: DYNJ49224I; 19) C SECTION PACK-LF, Model Number: DYNJ52387I; 20) C-SECTION PACK, Model Number: DYNJ56

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution

Worldwide distribution - US Nationwide.

Key facts

Status
Ongoing
Initiation date
2026-01-07
Report date
2026-02-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1417-2026