Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: D
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Worldwide distribution - US Nationwide.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1419-2026