FDA recall Z-1420-2018

Autochair · Class II · device

Product

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Reason for recall

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Distribution

PA, and worldwide

Key facts

Status
Ongoing
Initiation date
2018-03-07
Report date
2018-04-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Alfreton, United Kingdom

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1420-2018