FDA recall Z-1420-2020

Tosoh Bioscience Inc · Class II · device

Product

AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Reason for recall

A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.

Distribution

Distributed to AZ, NE, NJ, NY, and WA. Foreign distribution to Cayman Islands, Costa Rica, and Uruguay

Key facts

Status
Terminated
Initiation date
2020-01-10
Report date
2020-03-11
Termination date
2020-09-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Grove City, OH, United States