FDA recall Z-1420-2026

Medline Industries, LP · Class II · device

Product

Medline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104; 2) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104A; 3) KIT OPHTHALMOLOGY CORNEAL TRAN, Model Number: DYKMBNDL159; 4) KIT OPHTHALMOLOGY CATARACT, Model Number: DYKMBNDL180; 5) KIT OPHTHALMOLOGY EYE MUSCLE, Model Number: DYKMBNDL42; 6) KIT SURG ONC PORT INSERTION, Model Number: DYKMBNDL43B; 7) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84B; 8) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98; 9) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98A; 10) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98B; 11) MINOR EYE PACK, Model Number: DYNJ17219C; 12) PEDS EYE/MUSCLE PACK, Model Number: DYNJ38791M; 13) EYE PACK, Model Number: DYNJ41207C; 14) OPHTHALMIC PACK, Model Number: DYNJ42043I; 15) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852J; 16) CUSTOM EYE PACK HA-LF, Model Number: DYNJ45585C; 17) EYE PLASTIC PACK, Model Number: DYNJ47859B; 18) EYE PLASTIC

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution

Worldwide distribution - US Nationwide.

Key facts

Status
Ongoing
Initiation date
2026-01-07
Report date
2026-02-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1420-2026