FDA recall Z-1429-2020

Seca · Class II · device

Product

Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009

Reason for recall

It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.

Distribution

US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada

Key facts

Status
Ongoing
Initiation date
2019-06-19
Report date
2020-03-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Hamburg, N/A, Germany

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1429-2020