FDA recall Z-1439-2020

Vyaire Medical · Class II · device

Product

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

Reason for recall

The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.

Distribution

US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.

Key facts

Status
Terminated
Initiation date
2019-11-25
Report date
2020-03-11
Termination date
2023-08-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mettawa, IL, United States

Primary source