FDA recall Z-1464-2026

Medline Industries, LP · Class II · device

Product

Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

Reason for recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-01-08
Report date
2026-03-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States