Product

Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Medium Model/Catalog Number: REF# 9525 (Medium) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.

Reason for recall

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-02-24
Report date
2025-04-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cleveland, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1467-2025