FDA recall Z-1472-2026

Agiliti Health - Ellis · Class II · device

Product

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Reason for recall

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-01-21
Report date
2026-03-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Ellis, KS, United States

Primary source