FDA recall Z-1489-2021

Ortho-Clinical Diagnostics, INc. · Class II · device

Product

VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.

Reason for recall

Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV, WY and the countries of Australia, Bermuda, Brazil, Canada, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.

Key facts

Status
Terminated
Initiation date
2021-03-26
Report date
2021-05-05
Termination date
2023-05-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rochester, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1489-2021