FDA recall Z-1490-2022

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · Class I · device

Product

Crome Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Crome HF CRT-D, Model Numbers: DTPC2D4, DTPC2D1; b. Crome HF Quad CRT-D, Model Numbers: DTPC2QQ, DTPC2Q1

Reason for recall

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Distribution

Worldwide Distribution

Key facts

Status
Ongoing
Initiation date
2022-06-22
Report date
2022-08-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mounds View, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1490-2022