FDA recall Z-1493-2019

Abbott Laboratories Inc. (St Jude Medical) · Class II · device

Product

TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)

Reason for recall

In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.

Distribution

US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA