FDA recall Z-1497-2019

Abbott Gmbh & Co. KG · Class II · device

Product

Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Reason for recall

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

Distribution

Worldwide Distribution - US Nationwide to Oklahoma and Texas. Distributed worldwide to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, MADAGASCAR, MEXICO, NETHERLANDS, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UAE, UNITED KINGDOM, VIETNAM.

Key facts

Status
Terminated
Initiation date
2019-04-05
Report date
2019-06-05
Termination date
2022-09-28