FDA recall Z-1498-2020

LivaNova USA Inc. · Class II · device

Product

LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.

Reason for recall

The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.

Distribution

U.S. Nationwide distribution including Puerto Rico. This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected. O.U.S. - none

Key facts

Status
Terminated
Initiation date
2020-02-24
Report date
2020-03-25
Termination date
2022-08-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Arvada, CO, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1498-2020