FDA recall Z-1503-2023

Spacelabs Healthcare, Inc. · Class II · device

Product

Ultraview SL (UVSL) Command Module, Model 91496

Reason for recall

Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.

Distribution

US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.

Key facts

Status
Ongoing
Initiation date
2023-04-12
Report date
2023-05-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Snoqualmie, WA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1503-2023