FDA recall Z-1508-2020

Stryker Corporation · Class II · device

Product

Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electrocautery Probe, J Tip, 5MM 0250070443-PKG, StrykeFlow Electrocautery Probe, L Tip, 5MM 0250070444-PKG, StrykeFlow Electrocautery Probe, Ball Tip, 5MM 0250070445-PKG, StrykeFlow Electrocautery Probe, Needle Tip 5MM 0250070446-PKG, StrykeFlow Electrocautery Probe, Spoon Tip, 5MM

Reason for recall

Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.

Distribution

OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea

Key facts

Status
Terminated
Initiation date
2019-09-12
Report date
2020-03-25
Termination date
2022-10-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Jose, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1508-2020