FDA recall Z-1509-2024

Baxter Healthcare Corporation · Class II · device

Product

Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed under the following models: a) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,EU, Product Code 34BFST-2; b) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,UK, Product Code 34BFST-4; c) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,AU, Product Code 34BFST-6; d) PROBP 3400, BLUETOOTH, MOBILE, EU, Product Code 34BXST-2; e) PROBP 3400, BLUETOOTH, MOBILE, UK, Product Code 34BXST-4; f) PROBP 3400, BLUETOOTH, MOBILE, ZA, Product Code 34BXST-7; g) PROBP 3400, SUREBP, MOBILE, EU, Product Code 34XFST-2; h) PROBP 3400, SUREBP, MOBILE, UK, Product Code 34XFST-4; i) PROBP 3400, SUREBP, MOBILE, AU, Product Code 34XFST-6; j) PROBP 3400, SUREBP, MOBILE, ZA, Product Code 34XFST-7; k) PROBP 3400, MOBILE, EU, Product Code 34XXST-2; l) PROBP 3400, MOBILE, UK, Product Code 34XXST-4; m) PROBP 3400, MOBILE, AU, Product Code 34XXST-6; n) PROBP 3400, MOBILE, ZA, Product Code 34XXST-7; o) PROBP MOBILE STAND CORD ASSY EUROPE, Prod

Reason for recall

Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.

Distribution

Global, US: NY, IL

Key facts

Status
Ongoing
Initiation date
2024-03-26
Report date
2024-04-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1509-2024