FDA recall Z-1511-2022

GE Healthcare, LLC · Class II · device

Product

Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.

Reason for recall

As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

Distribution

US Nationwide distribution including PR. There was also government/military distribution. The countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mauritania, Mexico, Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, N

Key facts

Status
Ongoing
Initiation date
2022-06-02
Report date
2022-08-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1511-2022