# FDA recall Z-1517-2025

> **Beckman Coulter Inc.** · Class II · device recall initiated 2025-02-26.

## Product

DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes:  C63519 / 14987666545058  C63520 / 14987666545065    The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

## Reason for recall

Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the operator at the analyzer or automatically triggered by predefined rules in System Configuration, the calculated result will not be recalculated using the constituent tests  rerun results. The system will report only the calculated result using the initial constituent tests  results to the operator and LIS/Remisol. The issue was identified by Beckman Coulter internal engineering testing. The defect may cause a delay in reporting patient results.

## Distribution

Worldwide Distribution: U.S. (nationwide) to states of: AL, CA, CO, GA, ID, IL, IN, LA, MA, AI, MN, MS, MT, NC, NM, NV, NY, OK, OR, PA, PR, SC, TN, TX, WA, and WI; O.U.S. (foreign) to countries of: Australia, Belgium, Bosnia/Herzegovina, Canada, Croatia, Egypt, Ethiopia, France, Germany, Ghana, India, Italy, Republic of Korea, Lebanon, Libya, Malaysia, Netherlands, North Macedonia, Oman, Paraguay, Philippines, Poland, Qatar, Saudi Arabia, Seychelles, Slovakia, South Africa, Spain and Switzerland

## Key facts

- **Recall number:** Z-1517-2025
- **Recalling firm:** Beckman Coulter Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-02-26
- **Report date:** 2025-04-09

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Brea, CA, United States