FDA recall Z-1526-2021

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · Class II · device

Product

Implantable cardiac resynchronization therapy pacemakers (CRT-P) - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Percepta family: Percepta CRT-P MRI SureScan, Percepta Quad CRT-P MRI SureScan. 2. Serena family: Serena CRT-P MRI SureScan, Serena Quad CRT-P MRI SureScan. 3. Solara family: Solara CRT-P MRI SureScan, Solara Quad CRT-P MRI SureScan.

Reason for recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2021-04-06
Report date
2021-05-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mounds View, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1526-2021