FDA recall Z-1527-2020

Micro Therapeutics Inc, Dba Ev3 Neurovascular · Class I · device

Product

ev3 Pipeline Flex Embolization Device. For neurological endovascular use.

Reason for recall

Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.

Distribution

Worldwide distribution. US nationwide, Argentina, Australia, Austria, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Jordan, Kazakhstan, Korea, Lebanon, Mexico, Netherlands, Norway, Paraguay, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam

Key facts

Status
Terminated
Initiation date
2020-02-14
Report date
2020-04-01
Termination date
2024-06-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irvine, CA, United States

Primary source