ev3 Pipeline Flex Embolization Device with Shield Technology. Sold OUS. For neurological endovascular use.
Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.
Worldwide distribution. US nationwide, Argentina, Australia, Austria, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Jordan, Kazakhstan, Korea, Lebanon, Mexico, Netherlands, Norway, Paraguay, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1528-2020