FDA recall Z-1545-2026

Trividia Health, Inc. · Class I · device

Product

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter

Reason for recall

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Distribution

Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.

Key facts

Status
Ongoing
Initiation date
2026-02-06
Report date
2026-03-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Fort Lauderdale, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1545-2026