FDA recall Z-1556-2025

GE Medical Systems, LLC · Class II · device

Product

Venue Fit Models: R3 (Model Number H45303VFSC, H45303VFSW), R4 (Model Number H45314VFS, H45314VFSW), Venue Fit R5 system (Model Numbers H45335VFS, H45335VFSW), R3 Surgery for China, R3 PeriOP for China, R3 Expert for China, R3 Pro for China, R3 Performance for China, R3 Intervention for China diagnostic ultrasound systems

Reason for recall

The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-03-06
Report date
2025-04-23
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1556-2025