FDA recall Z-1559-2025

Baxter Healthcare Corporation · Class II · device

Product

Welch Allyn CONNEX Accessory Power Management Stand:

Reason for recall

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-03-24
Report date
2025-04-23
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1559-2025